Supplier: GE

This Panametrics aluminum oxide moisture sensor probemeasures moisture concentration in gases and
non-aqueous liquids from trace to ambient levels. It is
designed to be used in conjunction with all Panametrics
moisture analyzers for industries including:
• Petrochemical
• Natural gas
• Industrial gas
• Semiconductor
• Furnace gas/heat treating
• Power generation
• Air dryer
• Pharmaceutical
• Aerospace
Applications Features
• Intrinsically safe when used in conjunction with
certified electronics or zener barriers
• Superior sensitivity, response speed and calibration
stability
• Optional temperature sensor
• Calibrations traceable to National Institute of
Standards and Technology (NIST) or National
Physical Lab U.K. (NPL)
• Designed for in situ applications or in conjunction with
sample conditioning systems
• True absolute humidity sensor
• Wide dynamic range

M Series
Specifications
Moisture Probe
Intrinsic Safety
Intrinsically safe when connected
to a Panametrics Moisture Series analyzer, PM880
portable hygrometer or intrinsically safe barriers in
accordance with the user’s manual.
M Series moisture probe:
BAS01ATEX1096X
II 1 G EEx ia IIC T4 (-20°C to +80°C), and
CSA C US Class I, Division 1, Groups A,B,C&D T4,
LR44204-23
European Compliance
Complies with EMC Directive 89/336/EEC and
PED 97/23/EC for DN<25
Type
Aluminum oxide moisture sensor
Calibration
Each probe is individually computer calibrated against
known moisture concentrations, traceable to NIST or
NPL.
Dew/Frost Point Calibration Ranges
• Overall capability: 140°F to -166°F (60°C to -10°C)
by request
• Standard calibration range: 68°F to -112°F
(20°C to -0°C) with data to -166°F (-110°C)
• Ultralow calibration range: -58°F to -148°F
(-50°C to -00°C) with data to -166°F (-110°C)
Accuracy
• ±3.6°F (±2°C) in range of 140°F to -85°F (60°C to -5°C)
• ±5.4°F (±3°C) in range of -86°F to -166°F
(-66°C to 110°C)
Repeatability
• ±0.9°F (±0.5°C) in range of 140°F to -85°F
(60°C to -5°C)
• ±1.8°F (±1.0°C) in range of -86°F to -166°F
(-6°C to -10°C)
Temperature
• Operating: -166°F to 158°F (-10°C to 70°C)
• Storage: Maximum 158°F (70°C)
Operating Pressure (Dependent on Mount)
• Flange process connection (M1):
5 :Hg to 75 psig (6 bar)
• Threaded process connection (M2):
5 :Hg to 5000 psig (345 bar)
Flow Range
• Gases: Static to 10,000 cm/s linear velocity at 1 atm
• Liquids: Static to 10 cm/s linear velocity at density of
1 g/cc
Response Time
Less than five seconds for 63% step change in moisture
content in either wet-up or dry-down cycle
Consult factory for specifications of other moisture probes
available for special applications.
Input Voltage
1 VAC, 77 Hz
Impedance Range
50 kS to 2 MS depending on water vapor
Limited Warranty
• Calibration: Six months from delivery
• Materials and workmanship: One year from delivery
Optional Temperature Sensor
Type
Nonlinear negative temperature
coefficient (NTC) thermistor (resultant temperature
linearized by microprocessor)
Operating Range
-2°F to 158°F (-0°C to 70°C)
Accuracy
±0.9°F (±0.5°C) overall
Response Time (Maximum)
1 second in well-stirred oil or 10 seconds in still air for a
63% step change in increasing or decreasing
temperature



Traceability guaranteed. Every M Series Panametrics aluminum oxide data sheet ships with a certificate of conformance, packing list and warranty documentation. Ask our team for a datasheet or cross-reference.

Buying guide

Frequently asked questions

What is the lead time for M Series Panametrics aluminum oxide data sheet?
Stock items are dispatched within 24 hours of order confirmation. If M Series Panametrics aluminum oxide data sheet is not held in stock we source it through our approved supplier network, typically within 3–10 working days depending on condition and quantity. AOG requirements are prioritised through our 24/7 desk.
Is Panametrics supplied new or refurbished?
Both options are usually available. We confirm the exact condition, service history and documentation before dispatch, and every order ships with a certificate of conformance.
Do you ship Panametrics parts internationally?
Yes. We ship worldwide with export documentation, and we handle freight, customs paperwork and insurance so the part arrives at your site, workshop or warehouse without delays.
How do I confirm Panametrics fits my equipment?
Send us the part number together with your equipment model, serial number or nameplate photo. Our engineers cross-check the part against the OEM documentation and confirm compatibility before you order.